Product Information Update

Release of new UVV filters for EIT instruments

A new and improved UVV filter for the instruments UV PowerMap and UV Map Plus with V-filter has been released by EIT. The new V-filters ensure the UVV filter values to stay within the accepted +-10%, at biannual calibration. We therefore recommend all our customers to have the UVV filter changed in these types of instruments at next calibration and to have EIT instruments calibrated biannually in future.

1. All UV PowerMaps and UV Map Plus V-filter
instruments should be send for calibration
and UVV filter change as soon as possible.

2. PowerPuck and Uvicure Plus II with V-filter
keep their existing V-filters, but needs
calibration every six month in future.

Technical Issues
For further information regarding the UVV filter issue or technical issues, please contact us at calibration@efsen.dk or phone +45 45 65 02 60.

Calibration Planning
Please contact Julie Boesen at julie@efsen.dk for a quotation or to plan the calibration and V-filter change of your EIT instrument.

Please note that our calibration department is closed between Christmas and New Year and the next calibration will therefore be performed on Thursday January 5th 2012.